14 communications setup – Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 331

<5. EtherLCD>
5-80
IM 11B08A01-01E
5.4.14 Communications
Setup
1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5
Communications Setup
COM # 1
>
- Type DCS
- Protocol MODBUS
- Trans contents * * * * * * * * * *
- Recep contents * * * * * * * *
- Trans speed 9600bps
- Procedure * * * * * * * * * * * * * * * * *
- Parity
Odd
- Com mode RTU mode
- Scaling coef Real
Menu
COM#
F1
F2
F3
F4
F5
F6
Figure 5.88
Examples of communications setup screen
F1 (Menu):
Displays the Table menu screen.
F6 (COM #): Sets a communication port number.
• Modbus protocol can be set to both COM1 and COM2. However, GCCU and GCCU or
Y-Protocol (GC1000/GC8, GC6, GCHC) and Y-Protocol cannot be set to COM1 and COM2
respectively.
• Items that cannot be set due to types and protocols are displayed with * (asterisk).
Type NA
DCS
GCCU
Protocol
×
MODBUS
GC1000/GC8
GC6
GCHC
×
Trans contents
Ч
Ч
○
○
○
○
Recep contents
Ч
Ч
○
○
○
○
Trans speed
×
○
○
○
○
○
Procedure
Ч
Ч
○
○
○
○
Parity
×
○
○
○
○
○
Com mode
Ч
○
Ч
Ч
Ч
Ч
Scaling coef
Ч
○
Ч
Ч
Ч
Ч
For the Y-Protocol, refer to “ Serial Communication (Y-Protocol)” of page App.D-21
Display Item
Name
Set Lower Limit Upper Limit
Unit
Remarks
COM #
1
2
None, DCS, GCCU
Type
○ *
None, DCS, GCCU
Protocol
○ *
MODBUS, GC1000/GC8, GC6,
GCHC * Only when the type is
DCS
Trans
contents
○ *
Analyzed value, Alarm, All data
Recep
contents
○ *
None, Command, All data, GC8
type
Trans speed
○ *
1200 bps, 2400 bps, 4800 bps,
9600 bps, 19200 bps, 38400
bps
Procedure
○ *
Without procedure, With
procedure
Parity
○ *
None, Odd, Even
Com mode
○ *
RTU mode, ASCII mode
Scaling coef
○ *
Real, 9999, 65535
*:
The settings are valid after rebooting or turning off/on its power.
2nd Edition : May 11, 2012-00