User program delete screen (period script) – Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 347

<5. EtherLCD>
5-96
IM 11B08A01-01E
Display Item
Name
Set Lower Limit Upper Limit
Unit
Remarks
Period #
Period script Number
1
64
GCM Number
○
0
6
Auto Start
○ *
Not done (not executable)
Done (executable)
Exec Interval
○
Date
Hour
Minute
Second
Ana end
Interval (Date)
○
1
31
Date
When not selected by Cycle
Time Unit, please display with
asterisk. However, leave internal
data as they are.
Interval (Hour)
○
1
148
Hour
Interval
(Minute)
○
1
1440
Minute
Interval
(Second)
○
1
3600
Second
Exec SYS
Number
○
1
6
Only SYS numbers belonging to
relevant GCM can be selected.
When Ana end is not selected
in the Cycle Time Unit, display *
(asterisk).
* When Exec interval is Ana
end, the relevant period
script is executed at the time
of Ana end for the specifi ed
SYS number.
*:
The settings are valid after rebooting or turning off/on its power.
User Program Delete Screen (Period script)
Press F3 (Delete).
1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5
User Program Delete
Period # 1
>
- User Program Delete Not deleted
Menu
Main
Period#
F1
F2
F3
F4
F5
F6
Figure 5.108
Example of User Program Delete (Period script) screen
F1 (Menu):
Displays the Table menu screen.
F2 (Main):
Displays the User Programming screen (Period script).
F6 (Period #): Specifi es a Period script number.
• The operation accepts user level above C only.
• When Delete is selected in the User Program Delete, the relevant period script fi le is
deleted. When a script is deleted, the auto start of the script is automatically changed to NA.
Any period script in the status Enable cannot be deleted (an error message is displayed).
• When the execution status is Enable and the user program is about to be deleted, the error
message “No execution because of executing” is displayed.
• Even when the user program is deleted, the script fi le in the SD card is not deleted.
However, the script on the user program is deleted.
2nd Edition : May 11, 2012-00