B) elapsed time bar indicator – Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 214

<4. GC-HMI (touch panel)>
4-27
IM 11B08A01-01E
F0431.ai
Current ON/OFF status of
atmospheric pressure
balancing valve of GCM
Current stream valve
number of GCM
Figure 4.31
Example of GCM General Information Display Area
Method number indicator
Indicates the method number (1 to 6).
The indication is blank when the GCM is in manual or process - stop state.
Stream valve indicator
Indicates the stream valve status, either in terms of stream valve number or OFF.
Atmospheric balanced valve indicator
Indicates the status (ON or OFF) of the atmospheric pressure balancing valve. This indication is
hidden if an atmospheric pressure balancing valve not available.
(B) Elapsed time bar indicator
This area indicates the elapsed time of GSM and its subordinate SYS, as well as the ON/OFF
timings of stream valve and atmospheric pressure balancing valve.
F0432.ai
GCM elapsed time
Stream valve ON/OFF bar
SYS elapsed time bar
SYS interval division line
Atmospheric balanced valve ON/OFF bar
Figure 4.32
Display Example of Elapsed Time Bar
The GCM elapsed time is indicated in terms of the elapsed time (in seconds) since the beginning
of analysis, and the ratio (percentage) of elapsed time to the analysis interval.
When the GCM is in manual mode, the time axis is automatically adjusted, so that the elapsed
time is indicated in terms of the ratio (percentage) to the full scale.
2nd Edition : May 11, 2012-00
4