Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 54

<1. Overview>
1-27
IM 11B08A01-01E
A
Pressure regurator for utility gas (1 to 4)
Hydrogen kimiting unit
Inner wall of the
isothermal oven
Pressure gauge for
utility gas (1 to 4)
Pressure gauge for
carrier gas (1, 2)
A
A-A
Connection:
6 mm or
1/4 inch tube
Sample 2 (IN)
Sample 1 (OUT)
Carrier gas 1(except for H
2
)
Pressure regurator for carrier gas 1
Pressure regurator for carrier gas 2
Restrictor 1
Restrictor 2
Restrictor 3
Restrictor 4
Restrictor 5
Restrictor 6
Sample 2 (OUT)
Sample 1 (IN)
Combustion air
Carrier gas 2(except for H
2
)
Air output (Atomospheric balance valve 2)
Air output (Atomospheric balance valve 1)
Pressure regurator
for carrier gas (1, 2)
Figure 1.9
Pressure and fl ow control section of the isothermal oven
11/03/22
15:15:45
Operating Parameters (1/8)
Oven #1
Carrier gas #1
> - Carrier gas type
N2
- Carrier gas pressure 140.0 kPa
Menu
Menu
Menu
Menu Oven#
Gas#
F1
F2
F3
F4
F5
F6
11/03/22
15:15:45
Operating Parameters (2/8)
Oven #1
Utility gas #1
> -
Detector number
1-1
- Utility gas function
Burner fuel
- Utility gas type
H2
- Utility gas pressure 260.0 kPa
- Utility gas fl owrate
28
Menu
Menu
Menu
Menu Oven#
Gas#
F1
F2
F3
F4
F5
F6
2nd Edition : May 11, 2012-00
1