Calibration peak set – Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 326

<5. EtherLCD>
5-75
IM 11B08A01-01E
Display Item
Name
Set Lower Limit Upper Limit
Unit
Remarks
GCM #
Cal #
Cal Number
0
6
Cal stream
○ 0
31*
* Only stream numbers and cal
streams belonging to GCM can
be set. (Stream setup screen )
Cal times
○ 0
30
Cal times + Val before cal times
+ Val after cal times must not
exceed 31.
Auto cal
specifi cation
○
Not done, Done
ValStr #
before and
after cal
○ 0
31**
** Only stream numbers and Val
streams belonging to GCM can
be set. (Stream setup screen)
Val before
and after cal
times
○ 0
30
Cal times + Val before cal times
+ Val after cal times must not
exceed 31.
Calibration peak set
Press F2 (Peak).
1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5
Calibration Peak Setup
Peak # 1
>
- Peak name
- Execution Executed
- Measuring unit 1 ( )
- Measuring range 1.000
- Std conc
1.000
- Std area 950.000
- Calib factor 2.063
Cal
Peak#
F1
F2
F3
F4
F5
F6
Figure 5.83
Example of calibration peak set screen
F2 (Cal):
Displays the calibration set screen
F6 (Peak #): Specifi es a peak number.
• Displays the stream peak set in Cal stream.
2nd Edition : May 11, 2012-00