5) analyzer base sampling unit (gcsmp) – Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 58

<1. Overview>
1-31
IM 11B08A01-01E
11/03/22
15:15:45
Temperature Setup
Oven #1
- Oven max temp
145.0 °C
- Oven temp setpoint 130.0 °C
- LSV temp setpoint
1***** °C
> - FPD temp setpoint
30.0 °C
Menu
Menu
Menu
Menu Oven# Oven#
F1
F2
F3
F4
F5
F6
Figure 1.11
Example of the operation condition confi guration screen of the GC-HMI (touch panel)
EtherLCD
Table 1.1
The maximum number of parts for each oven
Large isothermal oven
Isothermal oven
Pressure regulator for
carrier gas
Max. 2
Max. 2
Restrictor
Max. 6
Max. 6 (4 recommended)
RV
LSV
Atmospheric-pressure
balancing valve
Max. 8 in total
Select from:
7 RVs, 1 LSV, 2 Atmospheric-pressure
balancing valves
Providing RV + LSV ≤ 7
Max. 7 in total
Select from:
5 RVs (3 recommended), 1 LSV, 2
Atmospheric-pressure balancing valves
Providing RV + LSV ≤ 5
Detector
Max. 2 in total
Select from:
2 TCDs, 2 FIDs, 1 FID-MC, 1 FPD
Max. 2 in total
Select from:
2 TCDs, 2 FIDs, 1 FID-MC
*FPD is not mountable.
Mesh arrester
Max. 2 (used for FID and FPD)
Flame arrester
Max. 5
(Connections for sample gas, carrier gas, utility gas, and I/O of the detector)
(5) Analyzer Base Sampling Unit (GCSMP)
In order to ensure the process gas chromatograph operates stably over a long period of time, it
is necessary to select the most appropriate sampling system corresponding to sample properties
in addition to analyzer stability and reliability. The GCSMP can be equipped on the self-standing
GC8000 Type 1 and 2, in which the analyzer and GCSMP can be operated and maintained
integrally.
The GCSMP adjusts the pressure and fl ow rate of the measurement stream for measuring the
process sample (gas and liquid) and the calibration and validation stream with standard gas
(liquid). The GCSMP also has a humidifying and warming function (setting temperatures: 50°C
or 75°C) and stream switching function by using air output from the analyzer. If the GCSMP is not
equipped, the stream switching valve can be operated by sending the air output or contact output
to an external sampling system.
2nd Edition : May 11, 2012-00
1