Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 69

<1. Overview>
1-42
IM 11B08A01-01E
● In the case where a stream sequence change command is received during the run mode in
the stream sequence status [1A2]
The measurement status is changed to the requested stream sequence status when the
measurement of the currently-measured stream is completed.
In this case, the operation mode remains unchanged from the run mode.
* Requests for changes to the same stream sequence number are invalid.
● In the case where a stream sequence change command is received during the pause mode
in the stream sequence status [1A6]
The measurement status is immediately changed to the requested stream sequence status.
In this case, the operation mode changes to the stop mode and valves are all closed (turned
off).
* Requests for changes to the same stream sequence number are invalid.
Stop
Run
W4
W5
W1
W2
W3
S2
S1
S5
S4
[1A1]
[1A2]
Time
Stream sequence 1
Stream sequence 2
Stream sequence 1
Assinning stream
sequence 2
Assinning stream sequence 1
Run command
Stop
Pause
Run
W4
W5
W1
W2
W3
S2
S1
S5
S4
[1A6]
[1A2]
Time
Stream sequence 1
Stream sequence 2
Stream sequence 1
Assinning stream
sequence 2
Assinning stream sequence 1
Run command
● In the case where a stream specifi cation change command (for continuous measurement)
is received during the run mode in the stream sequence status [1B2]
The measurement status is changed to the requested stream specifi cation status (for
continuous measurement) when the measurement of the currently-measured stream is
completed.
In this case, the operation mode remains unchanged from the run mode.
Run
W4
W1
W2
S2
S1
S4
W4
S4
W4
S4
W4
S4
W4
[1B2]
Time
Stream sequence 1
Stream specification 4 (Continuous)
Assinning stream specification 4 (Continous)
● In the case where a stream specifi cation change command (for N times of measurement) is
received during the run mode in the stream sequence status [1C2]
The measurement status is changed to the requested stream specifi cation status (for
N times of measurement) when the measurement of the currently-measured stream is
completed.
In this case, the operation mode changes from the run mode to the stop mode when the Nth
round of the specifi ed measurement is completed.
2nd Edition : May 11, 2012-00