Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 137

<2. Installation, Piping, and Wiring>
2-20
IM 11B08A01-01E
Hex.socket set screw
Cable packing adapter
Nut
Connection threads
Case of electronics section
(Pressurized enclosure)
Gasket
Body
Identification mark
Double-deck packing
(Replace it with an appropriate one from the accessory kit.)
Guide ring
Gland
Coupling
Cover
Cable
Hex key wrench
Hex.socket set screw
Plate
This plate is included with shipment.
Please keep the removed the plate.
To the connection port of
the electronics section
Figure 2.14
Procedure for mounting a cable packing adapter for the electronics section
The cable packing adapters comply with the Technical Standard of the Ministry of Health, Labour
and Welfare, Japan.
Table 2.7
Double-deck packing size for electronics section
Connection
port screw
Applicable
cable O.D.
Identifi cation mark
Recommended
torque for the
gland (N·m)
G3/4
ø9.0 to ø10.0
SCFP2010 ø9.0 to ш10.0
25
ш10.0 to ø11.0 SCFP2011 ø10.0 to ш11.0
20
ш11.0 to ø12.0 SCFP2012 ø11.0 to ш12.0
20
ш12.0 to ø13.0 SCFP2013 ø12.0 to ш13.0
20
ш13.0 to ø14.0 SCFP2014 ø13.0 to ш14.0
25
ш14.0 to ø15.0 SCFP2015 ø14.0 to ш15.0
20
ш15.0 to ø16.0 SCFP2016 ø15.0 to ø16.0
20
*:
When a cable O.D. falls under two categories, try both and choose the more suitable one.
For Ethernet cable, use sealing fi ttings in the accessories kit for the connection port (see Figure
2.15). Remove the attached cable packing adapter and mount the sealing fi tting back in place.
Six connection ports are provided in the electronics section. Use convenient ones.
2nd Edition : May 11, 2012-00