Detector type: fid or fpd – Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 288

<5. EtherLCD>
5-37
IM 11B08A01-01E
TIP
The Det operation, MC voltage value are refreshed at a one-second interval.
Display Item
Name
Set Lower Limit Upper Limit
Unit
Remarks
Oven #
Oven number
1
3
Det #
Detector number
1
2
Detector type
FID-MC
Det operation
○
0: Off
1: On
* For this operation, the user
level C or higher is required and
the operation mode must be
Stop in Manual status.
MC voltage
value
0.0
20.0
V
Target MC
voltage
○
Numeric, period “.” and slash “/”
are enabled: 16 characters
Input format: AA.A/B.B
Output format: AA.A+/–B.B mA
Detector type: FID or FPD
1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5
Det Adjustment
1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5
Det Adjustment
Oven #
1 Det # 1 ( FID )
Oven #
1 Det # 1 ( FPD )
>
- Det operation Off
>
- Det operation Off
Menu
Oven#
Det#
Menu
Oven#
Det#
F1
F2
F3
F4
F5
F6
F1
F2
F3
F4
F5
F6
Figure 5.48
Example of the detector adjustment screen (FID or FPD)
F1 (Menu):
Displays the Table Menu screen.
F5 (Oven #): Sets an oven number.
F6 (Det #):
Sets a detector number.
• The Det operation is refreshed at a one-second interval.
• The numbers of ovens to which no detectors are installed cannot be set via F5.
Display Item
Name
Set Lower Limit Upper Limit
Unit
Remarks
Oven #
Oven number
1
3
Det #
Detector number
1
2
Detector type
FID
FPD
Det operation
○
0: Off
1: On
* For this operation, the user
level C or higher is required and
the operation mode must be
Stop in Manual status.