Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 221

<4. GC-HMI (touch panel)>
4-34
IM 11B08A01-01E
Ref stream/Peak set
Set the reference stream number and reference peak number for the peak of the specifi ed
stream number, SYS number, and peak number.
Figure 4.42
Specifi c items to be set (peak 1 - ref stream number/peak 2 - ref stream number /peak 1 - ref
peak number/peak 2 - ref peak number) depend on the process assignment defi ned by stream
number/peak number.
Process assignment is external cubic: Peak 1 only
Process assignment is external linear: Peaks 1 and 2
Otherwise: No
peak
Setting range:
Peak 1 - Ref. stream number: 1 to 31
Peak 2 - Ref. stream number: 1 to 31
Peak 1 - Ref. peak number: 1 to 999 (only the peak number assigned to the reference stream
number can be selected)
Peak 2 - Ref. peak number: 1 to 999 (only the peak number assigned to the reference stream
number can be selected)
Gate /Integ. Set
Set the gate cut method and integration method for the peak of the specifi ed stream number,
SYS number, and peak number.
2nd Edition : May 11, 2012-00