Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 319

<5. EtherLCD>
5-68
IM 11B08A01-01E
Display Item
Name
Set Lower Limit Upper Limit
Unit
Remarks
Ref peak: Stream number
1
31
Available for “Calibration” or
“Calibration run”.
Ref peak: Peak number
1
999
Ref peak 1: Stream number
1
31
Ref peak 1: Peak number
1
999
Ref peak 2: Stream number
1
31
Ref peak 2: Peak number
1
999
Ref peak nume: Stream number
Ref peak nume: Peak number
Ref peak deno: Stream number
Ref peak deno: Peak number
Ref peak l: Stream number
Ref peak l: Peak number
Ref peak m: Stream number
Ref peak m: Peak number
○
1
1
1
1
1
1
1
1
31
999
31
999
31
999
31
999
Ratio (numerator)
Ratio (denominator)
Separation capacity (component l)
Separation capacity (component m)
Action time
○
0.0 or 5.0
21600.0 *
second * The upper limit value
depends on the sampling
rate of the detector.
Sampling rate Upper limit
20 ms
2700.0 s
40 ms
5400.0 s
80 ms
10800.0 s
160 ms
21600.0 s
• Minimum value is 5.0
seconds for noise level.
Std area
○
0.000
9999.999
Calib factor
○
0.000
9999.999
Area %
○
0.5
99.5
RVP coef A
○
-9999.999
9999.999
RVP coef B
○
-9999.999
9999.999
Std B.P
○
-999.9
999.9
Set up Conc.
0.00000
1.00000
Section set stop
○
0.0
0% value
○
-999.999
9999.999
A/I signal 1 to 16
100% value
○
-999.999
9999.999
H2 peak no
○
Compress fact.(air)
○
-9.9999
9.9999
Temp.(degC)
○
-9.9999
999.99
SYS number
○
1
6
• The SYS that belong to
the GCM that belong to the
stream is available.
• When execution is set to Executed, process assignment cannot be changed to No
processing.
1. When 0 is entered for both stream and peak number, synchronization is released.
2. Individual peak number cannot be set.
3. Setting synchronized peak of parent peak is invalid.
4. Cannot set peak within the same stream to synchronized destination.
5. Cannot set peak within calibration stream to synchronized destination.
6. Calibration run stream cannot be set for synchronization.
7. Only sample run streams can be set up for synchronization.
8. Distillation analysis stream and normal analysis stream can be synchronized.
2nd Edition : May 11, 2012-00