Detector type: fid-mc - reset screen – Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 286

<5. EtherLCD>
5-35
IM 11B08A01-01E
Display Item
Name
Set Lower Limit Upper Limit
Unit
Remarks
Oven #
Oven number
1
3
Det #
Detector number
1
2
Detector type
TCD
Det operation
○
Off, On
* For this operation, the user
level C or higher is required and
the operation mode must be
Stop in Manual status.
Voltage value
0.0
20.0
V
Current value
0.0
500.0
mA
Target current
○
Numeric, period “.” and slash “/”
are enabled: 16 characters
Input format: AA.A/B.B
Output format: AA.A+/–B.B mA
Detector type: FID-MC - Reset screen
Press F2 (Adjust).
1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5
Det Adjustment
Oven #
1 Det # 1 ( FID-MC )
>
- Det operation Off
- Reset status Initialized
Menu
Adjust
Reset
Oven#
Det#
F1
F2
F3
F4
F5
F6
Figure 5.46
Example of the detector type: FID-MC - reset screen
F1 (Menu):
Displays the Table Menu screen.
F2 (Adjust):
Displays the Det Adjustment (Execution) screen.
F4 (Reset):
Initializes a voltage value.
Pressing F4 displays the options “Abort” and “Initialize.” If the Set/Ent key
is pressed after the selection of “Initialize,” the MC voltage value is set to its
minimum value 10.0 V.
F5 (Oven #): Sets an oven number.
F6 (Det #):
Sets a detector number.
• Pressing F4 displays the options “Abort” and “Initialize” in the second line on the bottom. If
the Set/Ent key is pressed after the selection of “Initialize,” the MC voltage value is set to its
minimum value 10.0 V.
• The MC voltage value is read at a one-second interval, when, if the value is 0x0, the status
is refreshed with “Initialized”; otherwise Setting done.
• The Det operation is refreshed at a one-second interval.
• The numbers of ovens to which no detectors are installed cannot be set via F5.
2nd Edition : May 11, 2012-00
5