Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 300

<5. EtherLCD>
5-49
IM 11B08A01-01E
F1 (Menu):
Displays the Table Menu screen.
F5 (Oven #): Sets an oven number.
F6 (Det #):
Sets a detector number.
TIP
• The oven numbers set to Not provided on the Temperature Control Set cannot be selected.
• If the Detector type has been set to None on the Detector Set screen, all the options are shown as *
(asterisk).
• The initial screen displays Oven number 1 and Det number 1.
• When A/O chromatogram is set to None, items for Auto gain and Whole gain value are
displayed but an asterisk is given for set value.
• The initial screen displays the smallest oven number among those set in the isothermal
ovens on the Temperature Control Set.
• The initial screen displays Oven number 1 and Det number 1.
Display Item
Name
Set Lower Limit Upper Limit
Unit
Remarks
Oven #
Oven number
1
3
Det #
Detector number
1
2
Detector type
0) None
1) TCD
2) FID
3) FID-MC
4) FPD
Auto gain
○
0) None (default)
1) Individual gain
2) Whole gain *1
*1 Only for the A/O
chromatogram
Whole gain
value
○
0
15
0(default)
*1 Only for the A/O
chromatogram
Auto zero
○
0)Automatic feed (default)
1)Auto zero value
2)Disabled
* If the value is Automatic feed,
the value is set to the zero value
automatically obtained at the
auto zero time.
* If the value is Auto zero value,
the value is set to the auto zero
value at the start of the analysis.
Auto zero
value
○
−1000.0000 1000.0000
Former chromatogram standard
value
Auto zero
time
○
0
21600
If the value is 0.0 seconds, the
value is set to the zero value
automatically obtained at the
start of the analysis cycle.
* Upper Limit will use the
shortest analysis cycle for the
SYS analysis cycle that the
detector belongs to.
2nd Edition : May 11, 2012-00