Type 3 (self-standing), Overview – Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 35

<1. Overview>
1-8
IM 11B08A01-01E
M12 screw
25±5
*3: The four outer holes are used
for
installation.
The height of M12 screw
(prepared by user) is 25 ±5 mm
from the floor.
*4: Required clearance needed for wiring.
(580)
400
300
590
(1280)
11
00
20
430
580
410
410
410
375
250
350
435
A
2051
(1855)
40
500
(580)
(190)
(800)
27.5
300
(435)
Liquid-sample
valve (LSV) *1
Liquid-sample
valve (LSV) *1
Liquid-sample
valve (LSV) *1
Base
Standard
isothermal oven
Standard
isothermal oven
Standard
isothermal oven
Type 3 (Self-standing)
Weight: approx. 220 kg
Protection system A *1
Holes for installation
4-ø15 *3
*4
(435)
Operation panel *1
Grounding wiring
(100 Ω or less)
Connection to electronics section *2
Wiring connection *2
Air outlet for stream swiching valve
6 mm or 1/4” tube
Purging air inlet
Rc1/4 or 1/4NPT
Vent for purging air
Inlet/outlet for sample gas, carrier gases, etc.
6 mm or 1/4” tube
Unit: mm
Pressure gauge *1
Control unit
*1: It depends on specifications.
*2: Wiring connections are shown in
right
figures.
View A
Vent header *1
(connection
Rc1/2 or
1/2NPT)
(62)
28
28
35
b
(62)
36
36
125
125
For TIIS (wiring connection: G3/4)
For FM-X (wiring connection: 3/4NPT)
For FM-Y (wiring connection: 3/4NPT)
For ATEX, IECEx (wiring connection: M25x1.5)
Connection 3/4NPT
Connection
G3/4
Connection G3/4
Connection 3/4NPT
Connection M25x1.5
For ATEX, IECEx (wiring connection: G3/4 or 3/4NPT)
Cable gland
Sealing fitting
(as accessories)
(for Ethenet cable)
Connection 3/4NPT
Sealing fitting
(as accessories)
(for Ethenet cable)
Cable gland
Cable gland (as accessories)
For TIIS (wiring connection: 3/4NPT)
Connection
3/4NPT
Cable gland (as accessories)
Connection a
Connector (as accessories)
Connector (as accessories)
Connector
(as accessories)
Wiring
connection a b
G3/4
31
3/4NPT
29
(85)
(85)
36
c
Protection system B *1
Wiring
connectiona b c
G3/4
31
30
3/4NPT
29
28
2nd Edition : May 11, 2012-00