3 d/i status – Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 258

<5. EtherLCD>
5-7
IM 11B08A01-01E
5.2.3 D/I
Status
The D/I (contact input) status is displayed on the screen.
1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5
D/I Status
Slot # 1
DI# Name Value Processing name
1 a1 Open Stream sequence
2 Open No assignment
3 Open No assignment
4 Open No assignment
5 Open No assignment
6 Open No assignment
7 Open No assignment
8 Open No assignment
Menu
Table
Slot#
F1
F2
F3
F4
F5
F6
Figure 5.7
Example of the D/I status screen
F1 (Menu):
Displays the Status Menu screen.
F4 (Table):
Displays the D/I Setup screen.
F6 (Slot #):
Sets a slot number.
TIP
• The Value data are refreshed automatically at a fi xed interval (1 second).
• On the initial screen, the smallest slot number among those of the D/I cards inserted is displayed.
• For the slot number setting, only slot numbers of D/I or DI/O cards can be accepted.
• Up to CH3 are displayed for DI/O cards.
If no D/I cards are inserted, the Status Menu screen displays “Not load” on the bottom.
Display Item
Name
Set Lower Limit Upper Limit
Unit
Remarks
Slot #
Slot number
1
5
Absolute slot number
DI #
D/I contact number
1
8
Relative channel number
Name
D/I name
8 single-byte alphanumeric
characters
Status
Status
Open/Close
Processing
name
Processing name
No assignment, Stream
sequence, Stream, Range
change Cal/Val, Operation
mode change, Alarm process
2nd Edition : May 11, 2012-00