9 alarm status, Mac address screen – Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 266

<5. EtherLCD>
5-15
IM 11B08A01-01E
The MODBUS communication status with the registered GC is displayed together with the IP
address of the GC.
Display Item
Name
Set Lower Limit Upper Limit
Unit
Remarks
MODBUS
client
MODBUS
connection status
None
Connected
Disconnected
MODBUS IP address
MAC Address screen
To display the MAC address, press F6 (MAC).
1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5
Ethernet Status
MAC
MAC address A 00 : 00 : 00 : 00 : 00 : 00
MAC address B 00 : 00 : 00 : 00 : 00 : 00
Menu
MODBUS
TCP
F1
F2
F3
F4
F5
F6
Figure 5.17
Example of the MAC address screen
Display Item
Name
Set Lower Limit Upper Limit
Unit
Remarks
MAC address
A, MAC
address B
Addresses are entered in decimal digits using periods as delimiters. For example, the address
192.168.0.24 can be entered as is.
5.2.9 Alarm
Status
The detailed contents of the alarm can be indicated at the GC-HMI.(Refer to “4.7 Alarm Screen”)
Pressing the function key “F2” causes a transition to the alarm history screen.
TIP
• The screen for GCM#1 to 6 indicates the alarm to be occurred at each GC module. It for GCM#0 displays
the entire hardware alarm.
• The display can be scrolled vertically by the alarm row using the vertical scroll keys.
2nd Edition : May 11, 2012-00