Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 71

<1. Overview>
1-44
IM 11B08A01-01E
● In the case where a stop command is received during the pause mode in the stream
sequence status [1E6]
The operation mode is immediately changed to the stop mode in the stream sequence
status.
Run
[1E6]
Time
Stream sequence 1
Stop command
W1
W2
S1
* Pause status
Stop
Pause time
● In the case where a forced stop command is received during the run mode in the stream
sequence status [1F2]
The operation mode is immediately changed to the stop mode in the stream sequence
status.
Run
W1
W2
S2
S1
[1F2]
Time
Stream sequence 1
Forced stop command
Stop
● In the case where a forced stop command is received while the operation mode is
scheduled to be changed to the stop mode or pause mode in the stream sequence status
[1F3], [1F4]
The operation mode is immediately changed to the stop mode in the stream sequence
status.
Run
(Waiting for stop or pause)
W1
W2
S2
S1
[1F3] [1F4]
Time
Stream sequence 1
Forced stop command
Stop
2nd Edition : May 11, 2012-00