D. valve on/off setting, E. atm balanced valve on/off setting – Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 224

<4. GC-HMI (touch panel)>
4-37
IM 11B08A01-01E
Setting range:
Select ON or OFF. The stream valve number is in the range of 1 to 31.
TIP
• This operation is available if the user level is set to C or higher.
• Set the GCM status to manual.
• Only the stream valve number assigned to the active GCM is selectable.
d. Valve ON/OFF setting
Select the valve setting: ON/OFF timing.
TIP
• This operation is available if the user level is set to C or higher.
• Set the GCM status to manual.
• Only the valve assigned to the specifi ed SYS can be set here.
Figure 4.47
e. ATM balanced Valve ON/OFF setting
Select the status of atmospheric pressure balancing valve from ON or OFF.
TIP
• This operation is available if the user level is set to C or higher.
• Set the GCM status to manual.
• Only the atmospheric pressure balancing valve assigned to the active GCM is selectable.
2nd Edition : May 11, 2012-00