Warning, Caution, Xvii – Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 18: Operation

xvii
IM 11B08A01-01E
RUN
ALARM
POWER
SWITCH
OVERRIDE
POWER
PRESSURE
PURGING
OVERRIDE
LED (Green) of analysis "RUN"
LED (Red) of "ALARM"
LED (Green) of "POWER"
Electronic section
(Pressurized enclosure)
Door of electronic section
Operating Display
Electronics section (Pressurized enclosure)
Cover of electronic section
Door of isothermal oven
POWER LED (Yellow)
PRESSURE LED (Green)
PURGING LED (Red)
OVERRIDE LED (Yellow)
Status Display
Cover
Light sensor
Override switch
LABEL
Protection system
(Explosionproof
/ Flameproof enclosure)
FM-Y
TIIS
FM-X
ATEX
IECEx
Figure 5
Operation
WARNING
• When the door or cover of the pressurized enclosure is opened, use a gas detector to
check that the concentration of explosive gases in the ambient atmosphere is less than the
allowable limit.
• Before opening the door and cover of the electronics section, turn off the power and wait for
at least 25 minutes after obtaining the permission of the administrator.
• Electrostatic charge may cause an explosion hazard.
Avoid any actions that cause the generation of electrostatic charge, such as rubbing with a
dry cloth.
CAUTION
The isothermal oven is extremely hot after turning off the power immediately. Keep the protection
gas supplied for more than an hour after turning off the power. Keep hands away from the oven
components.
2nd Edition : May 11, 2012-00