App.g-18 – Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 517

App.G-18
IM 11B08A01-01E
The WAIT command cannot be used in the script commands for peak detection
stop time (due to possible delay in analysis result output).
Error will occur if the WAIT command is used in the script commands for peak
detection stop time.
WRITE_NV_DBL Command
Function
Writes the data in double precision real number into the element number specifi ed
by the double precision real number type nonvolatile memory.
Format WRITE_NV_DBL(N,
D)
N: Element number to read out data (0 to 255)
D: Write data (DOUBLE type)
Return value: None
Explanation When data type of write data differs from the command specifi cations, it will be a
syntax error.
(Example: When an integer variable is specifi ed in the second parameter of the
WRITE_NV_DBL command)
W’RITE_NV_SNG Command
Function
Writes the data in single precision real number into the element number specifi ed
by the single precision real number type nonvolatile memory.
Format WRITE_NV_SNG(N,
D)
N: Element number to read data (0 to 511)
D: Write data (REAL type)
Return value: None
Explanation See
WRITE_NV_DBL.
WRITE_NV_INT Command
Function
Writes data in integer into the element number specifi ed by the integer nonvolatile
memory.
Format WRITE_NV_INT(N,
D)
N: Element number to read out data (0 to 1023)
D: Write data (INTEGER type)
Return value: None
Explanation See
WRITE_NV_DBL.
2nd Edition : May 11, 2012-00