Peak copy 1 – Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 322

<5. EtherLCD>
5-71
IM 11B08A01-01E
Peak copy 1
Press F2 (Copy1).
1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5
Peak Copy 1
(1/4)
Total Assign # 30
Strm# Stream name Assign Peak num
1 A1 10 1
2 A2 10 5
3 10 5
4 0 0
5 0 0
6 0 0
7 0 0
8 0 0
Menu PeakSet
Copy
F1
F2
F3
F4
F5
F6
Figure 5.79
Example of peak copy 1 (1/4) screen
F1 (Menu):
Displays the Table Menu screen.
F2 (PeakSet): Displays the Peak Setup-Specifi c screen.
F6 (Copy):
Executes copy after entering Source stream and Destination stream.
• Copy destination peak must be an independent peak.
• See Peak synchronization and peak copy items regarding data for copy.
Display Item
Name
Set Lower Limit Upper Limit
Unit
Remarks
Total Assign # Total assign
0
999
Total assigned peak of all
streams
Strm#
Stream number
1
31
Stream name
Assign
0
999
Total assigned peak of stream
Peak num
0
999
Total used peak number of
stream
1. Copy peak assigned to stream source and destination.
2. When stream source and destination of assignment number are different, use the stream
with the smaller assignment number as the standard.
3. When the assignment number of either the stream source or destination is zero, error will
occur and copy cannot be made.
2nd Edition : May 11, 2012-00