Type 2 (wall-mounting), Overview – Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 32

<1. Overview>
1-5
IM 11B08A01-01E
(580)
400
300
590
11
00
(1280)
20
102
102
580
375
560
410
70 30
674
(80)
A
50
50
30
865
1541
(1345)
430
20
(70)
(800)
(190)
(580)
Liquid-sample
valve (LSV) *1
Liquid-sample
valve (LSV) *1
Large isothermal oven
Standard isothermal oven
Holes for installation
4-ø10.5
Type 2 (Wall-mounting)
Weight: approx. 155 kg
Protection system A *1
*6
*5
*4
(430)
*4: Required clearance needed for wiring.
*5: It is recommended to mount at approx. 500 mm
above the floor for easy operation or
maintenance.
*6: The wall construction for mounting has to be
designed for 4 times the weight of the analyzer.
Operation panel *1
Grounding wiring
(100 Ω or less)
Connection to electronics section *2
Wiring connection *2
Air outlet for stream swiching valve
6 mm or 1/4” tube
Purging air inlet
Rc1/4 or 1/4NPT
Vent for purging air
Inlet/outlet for sample gas, carrier gases, etc.
6 mm or 1/4” tube
Unit: mm
Pressure gauge *1
Control unit
*1: It depends on specifications.
*2: Wiring connections are shown in
right
figures.
View A
Vent header *1
(connection
Rc1/2 or
1/2NPT)
(62)
28
28
35
b
(62)
36
36
125
125
For TIIS (wiring connection: G3/4)
For FM-X (wiring connection: 3/4NPT)
For FM-Y (wiring connection: 3/4NPT)
For ATEX, IECEx (wiring connection: M25x1.5)
Connection 3/4NPT
Connection
G3/4
Connection G3/4
Connection 3/4NPT
Connection M25x1.5
For ATEX, IECEx (wiring connection: G3/4 or 3/4NPT)
Cable gland
Sealing fitting
(as accessories)
(for Ethenet cable)
Connection 3/4NPT
Sealing fitting
(as accessories)
(for Ethenet cable)
Cable gland
Cable gland (as accessories)
For TIIS (wiring connection: 3/4NPT)
Connection
3/4NPT
Cable gland (as accessories)
Connection a
Connector (as accessories)
Connector (as accessories)
Connector
(as accessories)
Wiring
connection a b
G3/4
31
3/4NPT
29
Protection system B *1
(only 100 V)
(85)
(85)
36
c
Wiring
connectiona b c
G3/4
31
30
3/4NPT
29
28
2nd Edition : May 11, 2012-00
1