6 creating reports, Creating reports -13 – Yokogawa Data Acquisition with PID Control CX2000 User Manual
Page 312
10-13
IM 04L31A01-01E
Computation and Report Function Related Operations
(Only on Models with the Computation Function Option)
10
10.6 Creating Reports
This section explains how to set the report type, report creation channels, report creation
time, and other items.
Measurement channels cannot be assigned to report channels on 0 measurement
channel models.
Procedure
Opening the Operation Display
Press the keys in the following sequence:
MENU key
(switch to Set mode) >
Hold down the FUNC key for 3 seconds
(switch
to basic setting mode) >
#6 soft key
(select [Option])
The following display appears.
Operations on the Display
1. Use the
arrow keys
to move the cursor (blue) to the item box you wish to change.
A soft key menu is displayed at the bottom of the display.
2. Press the
soft key
corresponding to the value you wish to select.
The box for the item you changed turns yellow, and the cursor moves to the next
item.
3. Repeat steps 1 and 2 to change the value of all the items you wish to change.
4. Press the
DISP/ENTER key
to confirm the changes.
The boxes for the items you changed turn from yellow to white, and the cursor
returns to the first item box.
Saving the Settings in Basic Setting Mode
1. Press the
ESC key
.
The display returns to basic setting menu.
2. Press the
[End] soft key
.
A confirmation dialog box appears
3. Select [Yes] and press the
DISP/ENTER key
.
The operation screen is displayed.