7-7
IM 04L31A01-01E
Measurement Function Related Setup Operations
7
• When setting the alarm relay output and alarm details
Press the keys in the following sequence:
MENU key
(switch to the Setting mode (Control)) >
Hold down the FUNC key for
3 seconds
(switch to basic setting mode) >
#1 soft key
(select [Alarm,A/
D,Temperature])
The following display appears.
• When setting the memory alarm
Press the keys in the following sequence:
MENU key
(switch to Set mode) >
Hold down the FUNC key for 3 seconds
(switch to basic setting mode) >
#9 soft key
(select [AUX, Time zone])
The following display appears.
Setup Procedure
1. Use the
arrow keys
to move the cursor (blue) to the item box you wish to change.
A soft key menu is displayed at the bottom of the display.
2. Press the
soft key
corresponding to the value you wish to select.
The box for the item you changed turns yellow, and the cursor moves to the next item.
3. Repeat steps 1 and 2 to change the value of all the items you wish to change.
Note
The value is set using the pop-up window that appears by pressing the [Input] soft key.
4. Press the
DISP/ENTER key
to confirm the changes.
The boxes for the items you changed turn from yellow to white, and the cursor
returns to the first item box.
Saving the Settings in Basic Setting Mode
1. Press the
ESC key
.
The display returns to basic setting menu.
2. Press the
[End] soft key
.
A confirmation dialog box appears.
3. Select [Yes] and press the
DISP/ENTER key
.
The operation screen is displayed.
7.2 Settings Related to Measurement Alarms