Yokogawa Data Acquisition with PID Control CX2000 User Manual

Page 155

background image

3-14

IM 04L31A01-01E

3.5

Initializing the Setup Data and Clearing the
Internal Memory

This section explains how to initialize the setup data in the internal memory to factory
default settings (initial settings) and how to clear the data in the internal memory. For a
list of initial settings, see appendix 6.

Procedure

Opening the Setting Display

Press the keys in the following sequence:

MENU key

(switch to Set mode) >

Hold down the FUNC key for 3 seconds

(switch to

basic setting mode) >

#5 soft key

(select [Load, Initialize]) >

#4

(select [Initialize])

The following setting display appears.

Setup Procedure

1. Use the

arrow keys

to move the cursor (blue) to the item box you wish to change.

The selections are displayed at the bottom section of the display.

2. Press the

soft key

corresponding to the value you wish to change.

The box for the item you changed turns yellow, and the cursor moves to the next item.

3. Repeat steps 1 and 2 to change the value of all the items you wish to change.
4. Press the

DISP/ENTER key

to confirm the changes.

The boxes for the items you changed turn from yellow to white, and the cursor
returns to the first item box.

Saving the Settings

1. Press the

ESC key

.

The display returns to basic setting menu.

2. Press the

[End] soft key

.

A confirmation dialog box appears.

3. Select [Yes] and press the

DISP/ENTER key

.

Setup Items

Select the type of operation from the following:
Clear1: Initializes the setup data of basic setting mode and setting mode and clears the

data in the internal memory.

Clear2: Initializes the setup data of setting mode and clears the data in the internal memory.
Clear3: Clears the data in the internal memory.
Data That Are Cleared
Items that are cleared are, display data, event data, manual sample data, TLOG data
(option), report data (option), and log information.