Caution, Location of the vga output terminal, Connecting to the monitor – Yokogawa Data Acquisition with PID Control CX2000 User Manual
Page 139
![background image](/manuals/360271/139/background.png)
2-22
IM 04L31A01-01E
2.5
Connecting a Monitor to the VGA Output
Terminal (/D5 Option)
CAUTION
• Connect the cable after turning OFF the CX2000 and the monitor.
• Do not short the VIDEO OUT terminal or apply external voltage to it. This may
cause damage to the CX2000.
Location of the VGA Output Terminal
The VGA output terminal is the D-sub connector labeled VIDEO OUT(VGA) at the upper
left corner of the rear panel of the CX2000.
!
VIDEO OUT(VGA)
Functions and Specifications of the VGA Output Terminal
The CX2000 display can be output to a monitor through RGB output. Monitors that can
be connected are VGA monitors or multi-sync monitors capable of displaying VGA.
Pin Assignments and Specifications of the VGA Output Terminal
Pin No. Signal Name
Specifications
1
Red
0.7 Vp-p
2
Green
0.7 Vp-p
3
Blue
0.7 Vp-p
4
–
5
–
6
GND
7
GND
8
GND
9
–
10
GND
11
–
12
–
13
Horizontal sync signal Approx. 31.5 kHz, TTL negative logic
14
Vertical sync signal
Approx. 60 Hz, TTL negative logic
15
–
Connecting to the Monitor
1. Turn OFF the CX2000 and the monitor.
2. Connect the CX2000 and the monitor using an analog RGB cable.
3. Turn ON the power to both the CX2000 and the monitor. The screen of the CX2000
is displayed on the monitor.
Note
•
When the CX2000 is turned ON, a VGA signal is always output from the VGA output
terminal.
•
The monitor screen may flicker if the CX2000 or another instrument is brought close to the
monitor.
•
The edge of the screen may drop out depending on the monitor type.
6
11
1
5
15
10
D-sub 15 pin receptacle