8 Manual control
viva 1.0 (for Process analysis)
■■■■■■■■
889
8.4.6.3.2
Digital output
Program part: Manual control
If a single digital output is selected in the device list, then the properties of
the digital output are displayed in the subwindow for functions/parame-
ters on the Switch on/off, Timer control and Set default tabs.
The status of the input is displayed in the subwindow for the graphical dis-
play.
Switch on/off
Port description
Display of the port description.
Port
Display of the internal description of the port.
Type
Display of the port type.
Parameter 1
The meaning of Parameter 1 depends on the Type of the digital output
that has been set.
When Type = Stirrer: stirrer output in %, pulse duration dependent on
Parameter 2.
Input range
-214783647 - 214783647
Default value
100
Parameter 2
The meaning of Parameter 2 depends on the Type of the digital output
that has been set.
For Type = Stirrer:
■
0: Duration of the switch-on pulse fixed at 500 ms.
■
>0: For 100% stirrer output (Parameter 1 = 100), pulse duration in
10 ms
Example:
Parameter 1 = 100, Parameter 2 = 30: Pulse duration switched-on
300 ms, switched off 0 ms
Parameter 1 = 75, Parameter 2 = 30: Pulse duration switched-on
225 ms, switched off 75 ms
Input range
-214783647 - 214783647
Default value
0