6 call cond, 1 call cond - overview, Call cond – Metrohm viva 1.0 (process analysis) User Manual
Page 431
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.0 (for Process analysis)
■■■■■■■■
419
Condition
on | off (Default value: off)
Expression for the condition(s) that can be created or edited with the for-
mula editor after pressing on
or by double-clicking in the text field
If this check box is activated, then the call is carried out only if the evalua-
tion of the condition yields the result 1 (= true).
Entry
1,000 characters
Default value
'empty'
5.5.2.4.6
CALL COND
5.5.2.4.6.1
CALL COND - Overview
Dialog window: Method
▶ CALL COND ▶ Properties... ▶ CALL COND - 'Command
name'
Command for the repeated calling of a VA track which continues to
be executed until one of the stop criteria is fulfilled. This command is used
chiefly for the conditioning of electrodes.
Appearance
The command has the following appearance:
Parameters
The parameters for the CALL COND command are configured in the fol-
lowing dialog window:
■
Command variables
The following command variables are generated by the CALL COND
command in the method run and can be used in formulas under the des-
ignation 'Command name.Variable identification':
Identification
Description
BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started