5 Method
viva 1.0 (for Process analysis)
■■■■■■■■
435
Selection
1 | off
Default value
1
off
Means that no stirrer will be used.
Stirring rate
Setting the stirring rate.
Input range
0 - 15
Default value
3
Switch off automatically
on | off (Default value: on)
If this check box is activated, the stirrer will be switched off automatically
when the command has finished.
5.5.2.5.4.3
MEAS Opt - Measuring parameters
Tab: Method
▶ MEAS Opt ▶ Properties... ▶ MEAS Opt - 'Command name' ▶ Mea-
suring parameters
Command name
Name of the command.
Entry
25 characters
The parameters for the measuring procedure can be defined on this tab.
Measuring parameters for the Avantes spectrometer
Wavelength
Wavelength at which the measured quantity (absorbance) is recorded.
Input range
100.0 - 2,000.0 nm
Default value
400.0 nm
Blank value
Value by which the measured quantity (absorbance) is corrected. Here an
expression can also be entered through the formula editor (e.g. xx.EME
from other measuring command) by which every measured absorbance
value can then be corrected.
Input range
-1E+12 - 1E+12 (max. 10 digits) mAU
Default value
0 mAU
Integration time
Integration time for absorbance measurement.
Input range
0.01 - 600,000 ms
Default value
6 ms