2 analog in - general – Metrohm viva 1.0 (process analysis) User Manual
Page 578
5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
566
■■■■■■■■
viva 1.0 (for Process analysis)
Identification
Description
BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
FIN
Command status; 1 = Command has ended at
least once, 0, invalid (variable not available) =
Command has never ended
UNI
Mapping unit which is defined for an IO port in the
configuration
[VAL]
Value of the command variables (optional, i.e.
'Command name.Portname' = 'Command
name.Portname.VAL'). The value returned corre-
sponds to the status of the IO port after the com-
mand is exited; 1 = active, 0 = inactive.
5.5.2.11.5.2
ANALOG IN - General
Tab: Method
▶ ANALOG IN ▶ Properties... ▶ ANALOG IN - 'Command
name'
▶ General
Command name
Name of the command.
Entry
25 characters
Device
Device name
Selection of a device from those available in the device table. Only devices
with which the command can be executed are available.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start
of the method.
Device type
Shows the instrument type.
Selection
IO Controller
Default value
IO Controller