6.8 Colorimetric sensors subwindow
■■■■■■■■■■■■■■■■■■■■■■
720
■■■■■■■■
viva 1.0 (for Process analysis)
Calibration data monitoring
on | off (Default value: off)
If this check box is activated, then the validity of the calibration is moni-
tored.
Calibration interval
Validity of the calibration in days. If a value is entered here, then the date
for the next calibration will be adapted automatically. The days are moni-
tored, i.e., the calibration becomes invalid when the date changes and not
at the time when the calibration took place.
Only editable for Calibration data monitoring = on, otherwise inactive.
Input range
0 - 999 days
Default value
999 days
Next calibration
Date on which the next calibration is to be carried out. The date can be
selected by clicking on
in the Select date dialog window. After the
date has been entered, the value for Calibration interval will be auto-
matically adjusted.
Only editable for Calibration data monitoring = on, otherwise inactive.
Default value Calibration date+ 999 days
Message
Message by e-mail
on | off (Default value: off)
The message is additionally sent to the address defined under [E-mail...]
if this check box is activated. The message is sent in text format.
Editable only if Calibration data monitoring = on, otherwise inactive.
[E-mail...]
[E-mail...] opens the Send e-mail window (see Chapter 2.5, page 70).
Active only if Message by e-mail = on.
Acoustic signal
on | off (Default value: off)
If this check box is activated, an acoustic signal will be emitted in addition
to the message.
Editable only if Calibration data monitoring = on, otherwise inactive.