5 evaluation - calibration, 1 calibration- overview, 2 calibration - general – Metrohm viva 1.0 (process analysis) User Manual
Page 601: 3 calibration - calibration curves, Evaluation - calibration, Calibration and
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.0 (for Process analysis)
■■■■■■■■
589
5.6.5
Evaluation - Calibration
5.6.5.1
Calibration- Overview
Subwindow section: Method
▶ Evaluation - Calibration ▶
Application
The way that the sample content for the substances is determined from
the peak data is defined in the Evaluation - Calibration section of the
evaluation. Only one calibration method can be selected for each method,
which is then valid for all available VA measuring commands.
Tabs
■
■
5.6.5.2
Calibration - General
Tab: Method
▶ Evaluation - Calibration ▶ General
This tab contains the basic settings for the calibration.
Calibration method
Selection of the method with which calibration is to take place.
Selection
no calibration | MLAT | DT
Default value
no calibration
MLAT
MLAT (Modified Linear Approximation Technique) is the default calibra-
tion method for the determination of brightener in electroplating
baths. Only the Linear regression curve type is permitted for this cali-
bration method (see Glossary, page 961).
DT
DT (Dilution Titration) is a calibration method for the determination of
suppressor in electroplating baths. The analysis is based on the fact
that galvanic metal precipitation is reduced by the addition of suppres-
sor to a VMS (Virgin Make-up Solution) (see Glossary, page 958).
5.6.5.3
Calibration - Calibration curves
Tab: Method
▶ Evaluation - Calibration ▶ Calibration curves
Selection of the measuring command
Measuring command
Selection list with all of the VA measuring commands with data acquisi-
tion defined in the method.
Selection
'VA measuring command'