2 cal spec - general/hardware, General/hardware – Metrohm viva 1.0 (process analysis) User Manual
Page 473
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.0 (for Process analysis)
■■■■■■■■
461
Identification
Description
BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
C0
Coefficient c0 of the calibration curve
C1
Coefficient c1 of the calibration curve
C2
Coefficient c2 of the calibration curve
C3
Coefficient c3 of the calibration curve
CAL
Status of the calibration: 1 = calibrated normally,
0 = canceled, because reference wavelength was
not found
CP{#}.MEA
Measured wavelength of the calibration point in nm.
If no measured peak can be assigned, then not
found will be entered
CP{#}.REF
Reference wavelength of the calibration point in nm
DBL
Total duration for the processing of the command in
s
FIN
Command status: 1 = Command has been ended
at least once, 0 = Command was started but
not ended, invalid = Command was not started
(variable not available)
STY
Type of stop with which the command was stopped:
1 = stopped normally, 0 = stopped manually
with [Quit] or [Stop]
TOL
Tolerance within which a measured peak can be
assigned to a calibration wavelength
5.5.2.6.2.2
CAL Spec - General/Hardware
Tab: Method
▶ CAL Spec ▶ Properties... ▶ CAL Spec - 'Command name' ▶ Gen-
eral/Hardware
Command name
Name of the command.
Entry
25 characters
The general parameters for the instrument are defined on this tab.