Metrohm viva 1.0 (process analysis) User Manual
Page 532
5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
520
■■■■■■■■
viva 1.0 (for Process analysis)
Device
Device name
Selection of a device from those available in the device table. Only devices
with which the command can be executed are available.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start
of the method.
Device type
Display or selection of the instrument type. If an instrument is selected in
the Device name field, the Device type field can no longer be edited
and the instrument type belonging to the instrument is displayed.
If the not defined option is selected in the Device name field, all instru-
ment types or groups with which the command can be executed can be
selected, regardless of the instruments present in the device table.
Selection
089 Photometer
Default value
089 Photometer
Heater
Target temperature
Temperature to which the cuvette has to be heated.
Input range
20 - 80 °C
Selection
Init. temp. | off
Default value
Init. temp.
Init. temp.
With Init. temp., the cuvette heater is activated at the same time as
the instrument and the cuvette is heated up to this temperature. The
initial temperature is entered and edited in the instrument configura-
tion (see Chapter 7.7.3, page 802).
off
Switch off heater.