3 properties - glp – Metrohm viva 1.0 (process analysis) User Manual
Page 705
■■■■■■■■■■■■■■■■■■■■■■
6 Configuration
viva 1.0 (for Process analysis)
■■■■■■■■
693
Message
Message by e-mail
on | off (Default value: off)
The message is sent to the address defined under [E-mail...] if this check
box is activated. The message is sent in text format.
[E-mail...]
[E-mail...] opens the Send e-mail window (see Chapter 2.5.1, page 70).
Acoustic signal
on | off (Default value: off)
If this check box is activated, an acoustic signal will be emitted in addition
to the message.
Action
If during monitoring it is found that the working life has expired then one
of the following actions will be triggered automatically during the start
test:
Selection
Record message | Display message | Cancel
determination
Default value
Display message
Record message
The message that the working life has expired will be automatically
saved with the determination.
Display message
A message is displayed and you can select whether you want to con-
tinue the run or cancel it. If the run is continued, the message that the
working life has expired is automatically saved with the determination.
Cancel determination
The ongoing determination will be automatically canceled. The follow-
ing message must be confirmed with [OK].
6.5.3.3
Properties - GLP
Tab: Configuration
▶ Devices ▶ [Edit] ▶ Properties... ▶ Properties - 'Device type' -
'Device name'
▶ GLP
Tab: Configuration
▶ Dosing units ▶ [Edit] ▶ Properties... ▶ Dosing unit - 'Na-
me'
▶ GLP
Tab: Configuration
▶ Solutions ▶ [Edit] ▶ Properties... ▶ Solution - 'Name' ▶ GLP
GLP test date
Date of the last GLP test. This date can be selected by clicking on
in
the Select date dialog window (see Chapter 2.4.1, page 68).