2 move - properties – Metrohm viva 1.0 (process analysis) User Manual
Page 514
5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
502
■■■■■■■■
viva 1.0 (for Process analysis)
Parameters
The parameters for the MOVE command are configured in the following
dialog window:
■
Command variables
The following command variables are generated by the MOVE command
in the method run and can be used in formulas under the designation
'Command name.Variable identification':
Identification
Description
BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
FIN
Command status; 1 = Command has ended at
least once, 0, invalid (variable not available) =
Command has never ended
RAN
Current absolute rotation angle of the rack in ° in
relation to the axis of the selected tower (entry when
exiting the command)
RPO
Current rack position (entry when exiting the com-
mand); 0 means 'not defined'
SAN
Current absolute swing angle of the robotic arm in °
(entry upon exiting the command)
5.5.2.8.2.2
MOVE - Properties
Dialog window: Method
▶ MOVE ▶ Properties... ▶ MOVE - 'Command name'
Command name
Name of the command.
Entry
25 characters
Device
Device name
Selection of a device from those available in the device table. Only devices
with which the command can be executed are available.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start
of the method.