5 Method
viva 1.0 (for Process analysis)
■■■■■■■■
525
814, 815, 858, 919
Selection
1 | 2 | 3
Default value
1
Set lines
Output signal
Entry of the bit pattern for the output signal or selection of a predefined
signal pattern.
It is possible to enter the following characters:
0 = Line inactive
1 = Line active
* = Any line status
p = Set pulse (pulse duration = 1 s). If a pulse with a different length is to
be output, then a corresponding template must be defined for this pur-
pose.
Sample changer, 894 and 846
Selection
Bit pattern with exactly 14 characters (0, 1, *,
p) | ************** | Signal pattern
Default value
**************
Avantes Spectrometer
Selection
Bit pattern with exactly 10 characters (0, 1, *,
p) | ********** | Signal pattern
Default value
**********
The bits are numbered from right to left:
Output 13 12 11 10 9 8 7 6 5 4 3 2 1 0
Bit
13 12 11 10 9 8 7 6 5 4 3 2 1 0
Example: Control of the 843 Pump Station via Remote Box
Output signal
Function
****p*********
Activating pump 1 for 1 s
***p**********
Activating pump 2 for 1 s
****1*********
Activating pump 1
***1**********
Activating pump 2
****0*********
Deactivating pump 1
***0**********