3 meas tmf - measuring parameters – Metrohm viva 1.0 (process analysis) User Manual
Page 467
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.0 (for Process analysis)
■■■■■■■■
455
Stirrer
Stirrer
Selection of the stirrer.
Selection
1 | off
Default value
1
off
Means that no stirrer will be used.
Stirring rate
Setting the stirring rate.
Input range
0 - 15
Default value
3
Switch off automatically
on | off (Default value: on)
If this check box is activated, the stirrer will be switched off automatically
when the command has finished.
5.5.2.5.6.3
MEAS TMF - Measuring parameters
Tab: Method
▶ MEAS TMF ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Entry
25 characters
Measurement
Two types of measurement can be selected from, Measurement with
drift control or Measurement without drift control.
Selection
Measurement with drift control | Measure-
ment without drift control
Default value
Measurement with drift control
Measurement with drift control
The measurement is carried out drift-controlled if this option is
selected. The measurement is canceled as soon as the defined Signal
drift or the Stop measured value is reached or the defined Max.
waiting time has expired. The corresponding parameters are not visi-
ble if this option is disabled.