13 start base point of the baseline, 14 end base point of the baseline – Metrohm viva 1.0 (process analysis) User Manual
Page 637
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.0 (for Process analysis)
■■■■■■■■
625
Description
Difference between the mean value of the peak heights of all replications
for the variation {x} and the mean value of the peak heights of all replica-
tions for the variation {x-1}.
Decimal places
0…2…6
Unit
#C
5.6.7.3.13
Start base point of the baseline
Variable name
RS.’Command name'.'Substance name'.VAR{x}.REP{y}.START
(with call-up of the VA measuring command by CALL VA)
RS.’Command name'.'Substance name'.VMS.REP{y}.START (with
call-up of the VA measuring command by CALL VMS)
RS.’Command name'.'Substance name'.INTERCEPT.REP{y}.START
(with call-up of the VA measuring command by CALL INTERCEPT)
RS.’Command name'.'Substance name'.'CALL
COND'{x}.REP{y}.START (with call-up of the VA measuring command
by CALL COND)
Description
Start base point, where the baseline of the peak evaluation begins.
Decimal places
0…3…6
Unit
V
5.6.7.3.14
End base point of the baseline
Variable name
RS.’Command name'.'Substance name'.VAR{x}.REP{y}.END (with
call-up of the VA measuring command by CALL VA)
RS.’Command name'.'Substance name'.VMS.REP{y}.END (with
call-up of the VA measuring command by CALL VMS)
RS.’Command name'.'Substance name'.INTERCEPT.REP{y}.END
(with call-up of the VA measuring command by CALL INTERCEPT)
RS.’Command name'.'Substance name'.'CALL
COND'{x}.REP{y}.END (with call-up of the VA measuring command by
CALL COND)
Description
End base point, where the peak evaluation baseline ends.
Decimal places
0…3…6
Unit
V