6 dosing - dosing a fixed volume, Fixed volume – Metrohm viva 1.0 (process analysis) User Manual
Page 884
8.4 Manual control - Functions
■■■■■■■■■■■■■■■■■■■■■■
872
■■■■■■■■
viva 1.0 (for Process analysis)
8.4.2.6
Dosing - Dosing a fixed volume
Program part: Manual control
The dosing of a preset volume can be started and stopped here. This tab
only appears if a single dosing device is selected.
Volume
Fixed volume that is to be dosed.
Input range
0.00 - 99,999.9 mL
Default value
1.000 mL
Dosing rate
The volume is dosed at this rate. The maximum dosing rate depends on
the cylinder volume of the dosing unit used. If the entered dosing rate is
too high for the selected dosing device, it will automatically be reduced
during dosing to the largest possible value.
NOTE
The dosing rate should be reduced for viscous liquids.
Selection
maximum
Default value
maximum
Input range
0.01 - 166.00 mL/min
Default value
4.00 mL/min
Filling rate
The buret will be filled at this rate after dosing. The maximum filling rate
depends on the cylinder volume of the dosing unit used. If the entered fill-
ing rate is too high for the selected dosing device, it will automatically be
reduced during filling to the largest possible value.
NOTE
The filling rate should be reduced for viscous liquids.
Selection
maximum
Default value
maximum
Input range
0.01 - 166.00 mL/min
Default value
4.00 mL/min