4 cal spec - calibration wavelengths, 5 cal spec - calibration wavelength, Calibration wavelengths – Metrohm viva 1.0 (process analysis) User Manual
Page 475
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.0 (for Process analysis)
■■■■■■■■
463
5.5.2.6.2.4
CAL Spec - Calibration wavelengths
Tab: Method
▶ CAL Spec ▶ Properties... ▶ CAL Spec - 'Command name' ▶ Cali-
bration wavelengths
Command name
Name of the command.
Entry
25 characters
The reference wavelengths with which the spectrometer is calibrated can
be entered on this tab.
Tolerance
Tolerance for the preset wavelength when assigning the highest peak
maximum found during the calibration measurement to the reference
wavelength. If a peak is outside the range Wavelength +/- Tolerance,
the calibration will be canceled. The existing calibration in the configura-
tion remains in effect.
Input range
0.1 - 1,100.0 nm
Default value
0.5 nm
[New]
Opens the Calibration wavelength # dialog window to enter wave-
lengths and tolerances (see Chapter 5.5.2.6.2.5, page 463).
[Properties]
Opens the Calibration wavelength # dialog window to edit the values
selected in the table (see Chapter 5.5.2.6.2.5, page 463).
[Delete]
Deletes the selected line.
5.5.2.6.2.5
CAL Spec - Calibration wavelength #
Tab: Method
▶ CAL Spec ▶ Properties... ▶ CAL Spec - 'Command name' ▶ Cali-
bration wavelengths
▶ [New] / [Properties] ▶ Calibration wavelength #
Wavelength
Input range
200.0 - 1,100.0 nm
Default value
'empty' nm