For Type = Stirrer:
■
0: Duration of the switch-on pulse set to 500 ms.
■
>0: With a stirring performance of 100% (Parameter 1 = 100), pulse
duration 10 ms
Example:
Parameter 1 = 100, Parameter 2 = 30: pulse duration switched on
300 ms, switched off 0 ms
Parameter 1 = 75, Parameter 2 = 30: pulse duration switched on
225 ms, switched off 75 ms
Input range
-214,783,647 - 214,783,647
Default value
0
7.11.6
Properties - Stepping motors
All recognized stepping motors are automatically listed on the Stepping
motors tab. The parameters for each stepper motor can be set.
Port
Shows the internal designation of the port.
Port description
Shows the port designation.
Type
Shows the port type.
Action
Shows the stepping motor action.
Distance/Angle
The meaning of the parameter depends on the Action set.
[Edit]
Opens the Properties stepping motor dialog in order to edit the
selected port.
[Port description]
Opens a menu in order to import or export port designations.
The [Port description] menu contains the following menu items:
Import
Import of a text file (*.txt; *.csv) containing the port designations. Each line is
interpreted as a port designation, empty lines will be ignored.
Export
Export of a text file containing the port designations of all digital inputs.