6 Configuration
viva 1.0 (for Process analysis)
■■■■■■■■
719
Concentration unit
Concentration unit for the lower or upper limit value of the calibration
range.
Selection
mol/L | mmol/L | µmol/mL | g/L | mg/L | µmol/L |
mg/mL | µg/mL | ppm | % | mEq/L
Default value
ppm
Calibration temperature
Temperature at which the calibration is carried out. In addition, the display
shows whether the temperature was measured with a Pt1000 or an NTC
temperature sensor or entered manually.
Input range
-20.0 - 150.0 °C
Default value
25.0 °C
NOTE
If the Calibration temperature has been entered manually, then the
following parameters will be reset to the default value or updated:
■
Calibration temperature = manual
■
Calibration date is updated.
■
Calibration method = manual
■
Determination ID is deleted
■
User is updated
Calibration date
Shows the date and time of the last calibration entered after each auto-
matic calibration or manual entry.
Calibration method
Display of the method name with which the last calibration has been car-
ried out. If the calibration data has been entered manually, then manual
will appear.
Determination ID
Identification of the determination.
The determination ID can be copied and pasted into another field.
User
Shows the short name of the user who was logged in during the calibra-
tion or who entered the calibration data manually. If the login function is
not used, then the user logged in under Windows will be entered auto-
matically.