Metrohm viva 1.0 (process analysis) User Manual
Page 411
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.0 (for Process analysis)
■■■■■■■■
399
12 Sweep
The sweep is executed in accordance with the defined sweep param-
eters.
13 Cleaning potential
Cleaning potential is applied to the electrodes.
14 Cleaning time
Idle until the period defined in the Cleaning time parameter has
expired.
only for hydrody-
namic measurement
= on
15 Stirrer off
The stirrer is switched off.
only for standby
potential = off
16 Potential off
The potential applied to the electrodes is switched off.
only for standby
potential
≠
off
17 Standby potential on
Standby potential is applied to the electrodes.
Parameters
The parameters for the CVS command are set on the following tabs:
■
General/Hardware (see Chapter 5.5.2.3.2.2, page 404)
■
Pre-treatment (see Chapter 5.5.2.3.2.3, page 406)
■
Sweep (see Chapter 5.5.2.3.2.4, page 407)
■
Post-treatment (see Chapter 5.5.2.3.2.5, page 409)
■
Potentiostat (see Chapter 5.5.2.3.2.6, page 410)
Command variables
The following command variables are generated by the CVS command in
the method run and can be used in formulas under the designation 'Com-
mand name.Variable identification':
Identification
Description
BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
DBL
Total duration for the processing of the command in
s