Metrohm viva 1.0 (process analysis) User Manual
Page 465
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.0 (for Process analysis)
■■■■■■■■
453
■
General/Hardware
Parameters for instrument.
■
Measuring parameters
Parameters for setting the measurement procedure.
Command variables
The following command variables are generated by the MEAS TMF com-
mand in the method run and can be used in formulas under the designa-
tion 'Command name.Variable identification':
Identifica-
tion
Description
BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command has
never been started
DBL
Total duration for the processing of the command in s
EME
End measured value (measured value after processing of
the command) in the unit of the measured value
ETE
End temperature (temperature following processing of
the command) in °C
FIN
Command status; 1 = Command has ended at least
once, 0, invalid (variable not available) = Command
has never ended
IME
Initial measured value (measured value before start con-
ditions are processed) in the unit of the measured value
ITE
Initial temperature (temperature before start conditions
are processed) in °C
LP.MEA
Measured value for the last measuring point of the mea-
suring point list in the unit of the measured value
LP.TEM
Temperature for the last measuring point of the measur-
ing point list in °C
LP.TIM
Time in s until the last measuring point in the measuring
point list is reached
MTE
Temperature measurement with sensor; 1 = on, 0 =
off)
NMP
Number of measuring points in measuring point list
STY
Type of stop with which the command was stopped: 1
= normal; 0 = manual or after error
TMF
Transmission factor