5 Method
viva 1.0 (for Process analysis)
■■■■■■■■
567
5.5.2.11.5.3
ANALOG IN - Stop criteria
Tab: Method
▶ ANALOG IN ▶ Properties... ▶ ANALOG IN - 'Command
name'
▶ Stop criteria
Command name
Name of the command.
Entry
25 characters
Stop criteria
Definition of possible stop criteria. The stop criterion that is reached first
leads to the cancellation of the command.
Maximum run time
on | off (Default value: off)
Check box for activation/deactivation of the stop criterion Maximum run
time. The command is canceled when the maximum run time has been
reached.
Input range
0 - 9,999.9
Default value
0
Selection
min | s
Default value
min
NOTE
The run time is also interrupted if the method is stopped with [HOLD].
Condition
on | off (Default value: off)
Check box for activation/deactivation of the stop criterion Condition.
Expression for the condition(s) that can be created or edited with the for-
mula editor after pressing on
or by double-clicking in the text field
The condition is checked continuously; if it is fulfilled, (evaluation of the
formula supplies the result 1 = true), then the command will be canceled.
Entry
1,000 characters
Default value
'empty'