9 rls dos, 1 rls dos - overview, Rls dos – Metrohm viva 1.0 (process analysis) User Manual
Page 510

5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
498
■■■■■■■■
viva 1.0 (for Process analysis)
846, 894
Selection
1 | 2 | 3 | 4
Default value
1
814, 815, 858, 919
Selection
1 | 2 | 3
Default value
1
089
Selection
1
Default value
1
Name of dosing unit
Selection of a dosing unit from the dosing units available in the table of
the dosing units.
Selection
'Name of the dosing unit' | not defined
Default value
not defined
'Name of the dosing unit'
Alphabetically organized selection of the names of the dosing units
from the table of dosing units in the configuration. A check is made
prior to the start of a method to determine whether the selected dos-
ing unit is identical with the name of the dosing unit at the selected
MSB connector.
not defined
No check is made prior to the start of a method to determine whether
the selected dosing unit is identical with the name of the dosing unit at
the selected MSB connector. This has the advantage that a method can
be utilized by users who use other dosing units.
5.5.2.7.9
RLS DOS
5.5.2.7.9.1
RLS DOS - Overview
Dialog window: Method
▶ RLS DOS ▶ Properties... ▶ RLS DOS - 'Command name'
Command to release a dosing device for using it in other parallel run-
ning methods. It removes the reservation made by the current method.
Instruments
This command can be executed with the following instruments:
Voltammetry: 894
Dosing Interface: 846
Sample changer: 814, 815, 858, 919
Photometer: 089