5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
434
■■■■■■■■
viva 1.0 (for Process analysis)
Device
Device name
Selection of a device from those available in the device table. Only devices
with which the command can be executed are available.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start
of the method.
Device type
Display or selection of the instrument type. If an instrument is selected in
the Device name field, the Device type field can no longer be edited
and the instrument type belonging to the instrument is displayed.
If the not defined option is selected in the Device name field, all instru-
ment types or groups with which the command can be executed can be
selected, regardless of the instruments present in the device table.
Selection
Avantes Spectrometer | 089 Photometer
Default value
Avantes Spectrometer
Sensor
This section is displayed only when Device type = 089 Photometer.
Temperature measurement
Type of temperature measurement
Selection
continuous | off
Default value
continuous
continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Measuring parameters tab will be
used.
Stirrer
This section is displayed only when Device type = 089 Photometer.
Stirrer
Selection of the stirrer.