Metrohm viva 1.0 (process analysis) User Manual
Page 787
■■■■■■■■■■■■■■■■■■■■■■
7 Devices
viva 1.0 (for Process analysis)
■■■■■■■■
775
NOTE
With the 858 Professional Sample Processor and the 919 IC Autosam-
pler plus, the properties of the tower, including the Swing Head
attached to it, are displayed on the Tower tab. In the case of the 814
USB Sample Processor and the 815 Robotic USB Sample Processor XL,
the model versions with one tower have one tab, Tower 1; model ver-
sions with two towers have two tabs, Tower 1 and Tower 2.
Tower parameters
Max. stroke path
Entry of the lowest permitted lift position. A lift height of 0 mm corre-
sponds to the "home position", i.e. the lift is moved all the way to the top.
Input range
0 - 235 mm
Default value
127 mm
Min. beaker radius
Definition of the minimum radius which the beakers used on the rack
must have. If the lift is moved to the work position, the beaker radii
defined in the rack table (see Chapter 6.11.2, page 743) for general sam-
ple positions (see Chapter 6.11.3, page 746) and special beakers (see
Chapter 6.11.3.2, page 746) are compared with the Min. beaker
radius. If this minimum beaker radius is not achieved, the run is stopped
and an error message is output. No check will be made if off is enabled.
Input range
1.0 - 100.0 mm
Selection
off
Default value
off
Lift rate
Lift rate for manual operation.
Input range
3 - 25 mm/s
Default value
25 mm/s
Axial distance
Distance between the axis of rotation of the sample rack and the swing
axis of the robotic arm.
814 USB Sample Processor, 858 Professional Sample Processor
Input range
100.0 - 300.0 mm
Default value
166.0 mm