5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
486
■■■■■■■■
viva 1.0 (for Process analysis)
This parameter can also be altered live after the first time the command
has been executed.
Input range
0.010 - 1,000.000 mL (Increment: 0.001)
Default value
0.010 mL
Addition
Selection of the addition mode for the sample.
Selection
Add manually | Add with dosing device
Default value
Add manually
Add manually
Manual addition of the sample, at which time the track is stopped. A
message appears and an acoustic signal is heard.
Add with dosing device
Automatic addition of the solution with a dosing device. The sample is
added through the Dosing port 1 defined in the configuration for the
dosing unit (see Chapter 6.6.3.2, page 699).
Message
With manual addition, the running of the track in which the command is
located is stopped and a message is displayed.
Only editable for Addition = Add manually, otherwise disabled (includ-
ing text input field).
Selection
Display standard message | Display message
defined by the user
Default value
Display standard message
Display standard message
If this check box is activated, then the standard message for additions
will be displayed.
Display message defined by the user
If this check box is activated, then the message defined in the text field
will be displayed.
Dosing device
Parameters for automatic dosing. This section is displayed only when
Addition = Add with dosing device.
Dosing unit
Selection of the dosing unit with which the sample solution is to be dosed
in.
Selection
'Name of the dosing unit' | not defined
Default value
not defined